| Oncology Systems - Low Energy Bunker Layout |
![]() |
|
|
Varian Low Energy
Load: Beam On - 12KVA
Voltage: (one of the following must be specified on order) 1. 208V/120 VAC, 70A, 60 Hz, 3Ø, 4 wire 2. 380V/220 VAC, 40A, 50 Hz, 3Ø, 5 wire
Branch circuit protection: A seperate grounding connector is required.
Variation: ±5% (maximum allowable deviation from nominal value selected) Installation per local code requirements is user's responsability.
Cooling Water Requirements: With incoming water temperature at 30°C (86°F), a flow of 11.4 liter/min (3 GPM) is required with a minimum pressure differential of 2 kg/cm²(30 PSI) at the machine base. The maximum overall pressure, shall not exceed 7 kg/cm² (100 PSIG). Incoming water temperature shall be between 10°C and 30°C (50°F-85°F). Power dissipation into cooling water is approximately 1 kW in ready mode and 7 kW at full power operation.
Ventilation Requirements: Room ventilation sufficient for removal of an air heat load of 5kW (17kBtu/hr) is required
|